When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]()
|
| Wrench, 1/4 inch for capillary connections |
1 | ![]() | 5043-0915 | Fitting mounting tool |
1 | ![]()
|
| Hexagonal key, 4 mm |
1 | ![]()
|
| Toothpick |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]()
|
| For parts, see |
Prerequisites
|
NOTE
The gaskets used in the standard and semi-micro/micro flow cell are different.
CAUTION
Sample degradation and contamination of the instrument
Metal parts in the flow path can interact with the bio-molecules in the sample leading to sample degradation and contamination.
Use a 4 mm hex key to unscrew the window assembly (1) and remove the gasket (6,7) from the cell body. Do not mix the gasket #6 and # 7.
Use a tooth pick to remove the quartz window from the window assembly.
NOTE
If the washers fall out of the window assembly, they must be inserted in the correct order with the PTFE ring to prevent any leaks from the flow cell window.
NOTE
Do not mix the gasket #6 and #7 (different hole diameter).
Using a 4-mm hex key, tighten the window screw hand tight plus a quarter turn.
Reconnect the capillaries, see Remove and Install a Flow Cell.
Perform a leak test.
Insert the flow cell.
Replace the front cover
Perform a Wavelength Verification Test or a Holmium Oxide Test to check the correct positioning of the lamp.