For bio-inert modules use bio-inert parts only! |
When
|
Tools required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]()
|
| Wrench, 1/4 inch |
1 | ![]() | 5043-0915 | Fitting mounting tool |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]() | G1315-60022 | Standard flow cell, 10 mm, 13 µL, 120 bar (12 MPa) |
1 | ![]() | G1315-60025 | Semi-micro flow cell, 6 mm, 5 µL, 120 bar (12 MPa) |
1 | ![]() | G1315-60024 | Micro flow cell, 3 mm, 2 µL, 120 bar (12 MPa) |
1 | ![]() | G1315-60015 | High pressure flow cell, 6 mm, 1.7 µL, 400 bar (40 MPa) |
1 | ![]()
|
| Nano flow cell |
1 | ![]() | G5615-60022 | Standard flow cell bio-inert, 10 mm, 13 µL, 120 bar (12 MPa) for MWD/DAD, includes 0890-1763 – 0.18 x 1500 mm PEEK capillary and 5063-6591 – PEEK fittings |
Prerequisites
|
CAUTION
Sample degradation and contamination of the instrument
Metal parts in the flow path can interact with the bio-molecules in the sample leading to sample degradation and contamination.
NOTE
The lamp house cover includes a magnet.
Disconnect the flow cell inlet capillary (top) from the union.
Disconnect the waste tubing (bottom) from the union.
Insert the flow cell capillaries into the union holder (top is inlet, bottom is outlet).
Reconnect the waste tubing (bottom) to the union.
Perform a Wavelength Verification and Calibration or a Holmium Oxide Test to check the correct positioning of the flow cell.