Error ID:
Dark-Current Test evaluation
Probable cause | Suggested actions | |
---|---|---|
1 | Defective slit assembly (stray light). | Run the Self-Test. Run the Self-Test. Run the Slit Test (G7117B) (part of the Self-Test). |
2 | Defective detector main board. | Please contact your Agilent service representative. |
3 | Defective PDA/optical unit. | Please contact your Agilent service representative. |
base-id: 3358540299
id: 3358540299-2