The slit test verifies correct operation of the micromechanical slit.
During the test, the slit is moved through all slit positions while the detector monitors the lamp intensity change. When the slit position is changed, the intensity drop (move to smaller slit) or intensity increase (move to larger slit) must be within a defined range.
If the intensity changes are outside the expected range, the test fails.
When
|
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]()
|
| Max-Light Cartridge Cell (filled with water) |
1 | ![]()
|
| Max-Light Cartridge Test Cell |
Prerequisites
|
Run the Slit Test with the Agilent Lab Advisor (for further information see Online-Help of user interface).