The self-test runs a series of individual tests (described on the next pages), and evaluates the results automatically. The following tests are run:
Slit Test (G7117B only)
Dark Current Test
Intensity Test
Wavelength Verification Test
ASTM Noise Test, a simplified version of the ASTM Drift and Noise Test (without testing the Drift)
When
|
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]()
|
| Max-Light Cartridge Cell (filled with water) , or |
1 | ![]()
|
| Max-Light Cartridge Test Cell |
Prerequisites
|
Run the Self-Test with Agilent Lab Advisor (for further information see Online-Help of user interface).
Under the tab Signals you can find the detailed signals from the tests.
base-id: 3526291723