Tools required
Qty. | p/n | Description | |
---|---|---|---|
2 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]()
|
| Column (up to 30 cm length) |
1 | ![]() | G7116-68003 | Column Holder Lamella, 2/pk |
For more information on recommended capillary types, refer to Capillary Connections.
Prerequisites
|
NOTE
A maximum of two columns can be installed.
Turn the column onto the Quick Connect Fitting until the first resistance is observed.
Fingertight
Route the capillary between the flap door and the ICC body towards the detector. Alternatively, route the capillary through the incisions at the right side of the ICC.
If applicable, connect the column identification tag to the tag reader, see Connect a Column Identification Tag to the Tag Reader.
Configure the connected column in the CDS, see Column Assignment.