Clicking on Column Assignment in the control interface opens a dialog box, where the column information can be edited:
Plumbing: The user can specify whether the column is located in the Upper or Lower position of the ICC heat exchanger, or indicate that no column is present in the ICC at all (None).
Visualization: This section gives a visual representation of the column configuration in accordance with the information in the Plumbing section.
Column Tag Information: This section holds information on the physical properties of the columns installed in the ICC and their usage (see Column Tag Information Table for more information). The fields of the table can be edited and the information therein used for reporting purposes even if no column identification tags are present. This is often referred to as soft tagging. When the Column ID Upgrade Kit is installed and a column identification tag is present, the content of the table corresponds to the information stored on the tag. This is referred to as hard tagging.
NOTE
By default, the Column Tag Information table is truncated and only columns up to and including Injections are displayed. Clicking on >> in the top right corner displays the whole table.