The intensity test measures the intensity of the UV-lamp over the full wavelength range (190 - 950 nm). Four spectral ranges are used to evaluate the intensity spectrum.
The test is used to determine the performance of the lamp, the cell, and the optical unit (see also Cell Test). When the test is started, the 1 nm slit is moved into the light path automatically.
To eliminate effects due to absorbing solvents, the test should be done with the flow cell removed. The shape of the intensity spectrum is primarily dependent on the lamp, grating, and diode array characteristics. Therefore, intensity spectra will differ slightly between instruments.
When
|
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]()
|
| Removed Flow Cell |
Prerequisites
|
Run the Intensity Test with Agilent Lab Advisor (for further information see Online-Help of user interface).
base-id: 3358600331
id: 9007202613341323