The cell test measures the intensity of the UV- and tungsten lamps over the full wavelength range (190 - 950 nm), once with the flow cell installed, and once with the flow cell removed. The resulting intensity ratio is a measure of the amount of light absorbed by the flow cell. The test can be used to check for dirty or contaminated flow cell windows. When the test is started, the 1 nm slit is moved into the light path automatically.
This test should be performed initially with a new detector/flow cell. The values should be kept for later reference/comparison.
When
|
Parts required
Qty. | p/n | Description | |
---|---|---|---|
1 | ![]()
|
| Flow Cell (filled with water) |
Prerequisites
|
Run the Cell-Test with Agilent Lab Advisor (for further information see Online-Help of user interface).