Intensity Test Evaluation
NOTE
If only one range fails and the application does not require this range, the lamp may not be changed.
Probable cause | Suggested actions | |
---|---|---|
1 | Incorrect calibration | Recalibrate and repeat the test. |
2 | Dirty or contaminated optical components. | Please contact your Agilent service representative. |
3 | Old UV-lamp. | Exchange the UV lamp. |
4 | Defect optical unit. | If the test fails with flow cell and new UV-lamp, please contact your Agilent service representative. If the test fails with new UV-lamp, please contact your Agilent service representative. |