Type | Specification | Comments |
---|---|---|
Weight | 14.5 kg (32 lbs) |
|
Dimensions (height × width × depth) | 180 x 396 x 436 mm (7.1 x 15.6 x 17.2 inches) |
|
Line voltage | 100–240 V~, ±10% | Wide-ranging capability |
Line frequency | 50 or 60 Hz, ±5% |
|
Power consumption | 80 VA, 65 W |
|
Ambient operating temperature | 4–55 °C (39–131 °F) |
|
Ambient non-operating temperature | -40–70 °C (-40–158 °F) |
|
Humidity | < 95% r.h. at 40 °C (104 °F) | Non-condensing |
Operating altitude | Up to 3000 m (9842 ft) |
|
Non-operating altitude | Up to 4600 m (15092 ft) | For storing the module |
Safety standards: IEC, EN, CSA, UL | Overvoltage category II, Pollution degree 2 | For indoor use only |
ISM Classification | ISM Group 1 Class B | According to CISPR 11 |
Type | Specification | Comments |
---|---|---|
Hydraulic system | Dual piston in series pump with proprietary servo-controlled variable stroke drive, floating pistons and active inlet valve, integrated 4-channel degassing unit |
|
Designed for use with Agilent InfinityLab Assist | Intuitive User Interface, Automated Workflows, Predictive Maintenance & Assisted Troubleshooting | |
Flow range | Settable: 0.001 – 10 mL/min Recommended: 0.2 – 10.0 mL/min | Set points in 0.001 mL/min increments |
Flow precision | ≤ 0.07 % RSD, or ≤0.02 min SD, whichever is greater | Based on retention time at constant temperature |
Flow accuracy | ± 1 % or 10 µL/min, whichever is greater | Pumping degassed H₂O at 10 MPa (100 bar, 1450 psi) |
Pressure operating range | Up to 60 MPa (600 bar, 8700 psi) up to 5 mL/min Up to 20 MPa (200 bar, 2950 psi) up to 10 mL/min |
|
Pressure pulsation | < 2 % amplitude (typically < 1.3 %), or < 0.3 MPa (3 bar, 44 psi), whichever is greater | At 1 mL/min isopropanol, at all pressures > 1 MPa (10 bar, 145 psi) |
Compressibility compensation | User-selectable, based on mobile phase compressibility | |
Recommended pH range | 1.0 – 13, short term 14 | |
Gradient formation | Low pressure quaternary mixing/gradient capability using proprietary high-speed proportioning valve | |
Delay volume | 600 – 900 µL, dependent on back pressure | Measured with water at 1 mL/min (water/water with tracer) |
Settable composition range | 0 – 100 % in 0.1 % increments | |
Composition precision | < 0.2 % RSD or < 0.04 min SD, whichever is greater | At 1 mL/min; based on retention time at constant temperature |
Instrument Control | LC & CE Drivers A.02.14 or above Lab Advisor software B.02.08 or above InfinityLab Assist (G7180A) with firmware D.07.40 or above | For details about supported software versions refer to the compatibility matrix of the LC & CE Drivers version in use |
Local Control | Agilent Instant Pilot (G4208A) B.02.20 or above | |
Communication | Controller-area network (CAN), USB Extended Remote Interface: ready, start, stop and shut-down signals |
|
Materials in flowpath | Titanium, Gold, Platinum-Iridium, Sapphire, PEEK, PTFE, Ruby, Ceramic, FEP, PFA |
|
Active seal wash | Included |
|
base-id: 4290901131