The items in the pump graphic have the following meaning and function:
| The level of solvent in the bottle is denoted by the gray area. When the solvent level falls below the specified volume, the area turns yellow. When the bottle is empty, the area turns red. Clicking the solvent bottle displays the Bottle Fillings dialog box. The tooltip for the bottle shows the solvent name. |
|
Indicates that the ISET option is installed but with no active method (light gray) or installed and active (dark gray). |
|
The pressure setpoints. The red line shows the current maximum pressure limit; the gray area shows the current pressure (also shown as text). |
The current solvent flow rate (in mL/min) is displayed above the pressure display.
The instrument signals are displayed when clicking the icon in the upper right corner of the module dashboard. The following pump signals are displayed:
Flow | The current solvent flow rate (in mL/min). |
Pressure | The current pump pressure (in bar, psi or MPa, see Instrument Configuration). |
Tuning |
Indicates the tuning efforts of the pump. For pumps operating as expected, the signal should stay in a range of –1 to +1 within the full-scale of –2 to +2. |
Pressure Limit | The current maximum pressure limit. |
Composition A:B | The contributions of channels A and B to the current solvent composition. |
Composition C:D |
The contributions of channels C and D to the current solvent composition. |
Valve position |
The current valve position. |
Prepare Pump | The status of a prepare pump task (purge, prime, or condition). |
base-id: 4133937419
id: 4133937419