Use the Instrument Configuration dialog box to examine and, if necessary, modify your instrument configuration. The Configurable Modules panel contains a list of all modules available for configuration. The Selected Modules panel contains the list of configured modules.
Auto Configuration: Under Communication settings, select either the Host Name option or the IP address option and enter the appropriate value for the host computer to enable automatic detection of the hardware configuration. The system configures the instrument automatically with no further manual configuration necessary.
The Quaternary Pump configuration parameters are in two sections:
Communication
Options
Communication: The parameters in this dialog box are detected automatically during autoconfiguration.
Device name,
Type ID,
Serial number,
Firmware revision,
Connection settings
Options:
Pressure Unit:
Select the pressure units from the drop-down list (bar, psi, or MPa).
Seal wash installed:
This check box is marked to indicate that an optional seal wash has been detected during autoconfiguration.
Installed mixer:
The installed mixer is detected during autoconfiguration. For manual configuration, click the down-arrow and select the installed mixer from the list or choose No mixer installed.
Configure Solvent Type Catalogs: Displays the Solvent Type Catalogs dialog box, which allows you to import and export solvent calibration data.
See How to Import the Latest Solvent Calibration Tables.
Please see the online help of your user interface for more detailed information.