错误 ID: 29203
The maximum stroke is too short.
During initialization the pump defines the operation position of the pump drives and therefore the pistons. First the pump drive moves backwards to find a mechanical stop within the ball screw. Afterwards, pistons move forwards for finding the maximum available stroke volume. These values are expected within a pre-defined range. "Maximum stroke too short" means that the outer drive position is too close. This can be caused by a drive initialization without pump head or if the pump head has not been installed properly (screws are loose).
Parameter: 1 – 2 referring to pump drive
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Wiper shifted | 请与您的 Agilent 服务代表联系。 |
2 | Pump head blocks piston movement | Replace, clean or repair pump head. |
3 | Pump drive motor is mechanically blocked. | 请与您的 Agilent 服务代表联系。 |
base-id: 4133183371