错误 ID: 2046, 2050, 2510
Encoder Missing: Pump channel A
B: Encoder Missing: Pump channel B
The optical encoder on the pump motor in the module is missing or defective.
The processor checks the presence of the pump encoder connector every 2 s. If the connector is not detected by the processor, the error message is generated.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Defective or disconnected pump encoder connector. | 请与您的 Agilent 服务代表联系。 |
2 | Defective pump drive assembly. | 请与您的 Agilent 服务代表联系。 |
base-id: 3944517643