错误 ID: 2243, 8016
The pump board gets no or wrong pressure signals from the built-in degasser.
Parameters:
0: ADC failure
1: ADC cannot convert pressure signal
2: Measured pressure out of range (negative or higher than 1021 hPa)
The pump board gets no or wrong pressure signals from the built-in degasser:
No valid pressure signal is shown during startup of the degasser
Measured pressure is higher than 32000 hPa
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Degasser board defective, missing, or not connected to the pump mainboard. | 请与您的 Agilent 服务代表联系。 |
2 | Degasser sensor defect or not connected to degasser board | 请与您的 Agilent 服务代表联系。 |
3 | If parameter is 0 and the LED on the degasser board blinks, it is probably because of a defective pump mainboard. | 请与您的 Agilent 服务代表联系。 |
4 | Degasser sensor defect. | 请与您的 Agilent 服务代表联系。 |
5 | Degasser sensor not connected to mainboard. | 请与您的 Agilent 服务代表联系。 |