错误 ID: 2040
(MCGV/4CGV)
Valve 0 Failed: valve A
Valve 1 Failed: valve B
Valve 2 Failed: valve C
Valve 3 Failed: valve D
One of the valves of the multi-channel gradient valve has failed to switch correctly.
The processor monitors the valve voltage before and after each switching cycle. If the voltages are outside expected limits, the error message is generated.
Valve 0 Failed: valve A1
Valve 1 Failed: valve A2
Valve 2 Failed: valve B2
Valve 3 Failed: valve B1
One of the solvent selection valves in the module failed to switch correctly.
The processor monitors the valve voltage before and after each switching cycle. If the voltages are outside expected limits, the error message is generated.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Gradient valve disconnected. | Ensure the gradient valve is connected correctly. |
2 | Solvent selection valve disconnected. | 请与您的 Agilent 服务代表联系。 |
3 | Connection cable (inside instrument) not connected. | 请与您的 Agilent 服务代表联系。 |
4 | Connection cable (inside instrument) defective. | 请与您的 Agilent 服务代表联系。 |
5 | Gradient valve defective. | Exchange the gradient valve. |
6 | Solvent selection valve defective. | Exchange the solvent selection valve. |
base-id: 3821272971