错误 ID: 2207, 2217, 25120
Initialization Failed: Pump channel A
B: Initialization Failed: Pump channel B
The module failed to initialize successfully within the maximum time window.
A maximum time is assigned for the complete pump-initialization cycle. If the time is exceeded before initialization is complete, the error message is generated.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Blocked (passive or active) inlet valve. | Exchange the (passive or active) inlet valve. |
2 | Defective pump drive assembly. | 请与您的 Agilent 服务代表联系。 |
3 | 主板出现故障。 | 请与您的 Agilent 服务代表联系。 |
base-id: 3821268491