错误 ID: 2205, 2215, 2505
Index Missing: Pump channel A
B: Index Missing: Pump channel B
The encoder index position in the module was not found during initialization.
During initialization, the first piston is moved to the mechanical stop. After reaching the mechanical stop, the piston reverses direction until the encoder index position is reached. If the index position is not recognized within a defined time, the error message is generated.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Disconnected or defective encoder cable. | 请与您的 Agilent 服务代表联系。 |
2 | Defective pump drive assembly. | 请与您的 Agilent 服务代表联系。 |
base-id: 3821051787
id: 9007203075792779