错误 ID: 2044
The active-inlet valve in the module has drawn excessive current causing the inlet-valve electronic fuse to open.
Inlet-Valve Fuse 0: Pump channel A
Inlet-Valve Fuse 1: Pump channel B
One of the active-inlet valves in the module has drawn excessive current causing the inlet-valve electronic fuse to open.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Defective active inlet valve. | Restart the module. If the error message appears again, exchange the active inlet valve. |
2 | Defective connection cable (front panel to mainboard). | 请与您的 Agilent 服务代表联系。 |
3 | 主板出现故障。 | 请与您的 Agilent 服务代表联系。 |
base-id: 3820966283