错误 ID: 2016
The pressure signal is missing.
The pressure signal must be within a specific voltage range. If the pressure signal is missing, the processor detects a voltage of approximately -120 mV across the damper connector.
The pressure signal must be within a specific voltage range. If the pressure signal is missing, the processor detects a voltage of approximately -120 mV across the pressure sensor.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Damper disconnected. | 请与您的 Agilent 服务代表联系。 |
2 | Pressure sensor disconnected. | 请与您的 Agilent 服务代表联系。 |
3 | 阻尼器出现故障。 | 请与您的 Agilent 服务代表联系。 |
4 | Defective pressure sensor. | 请与您的 Agilent 服务代表联系。 |
base-id: 3820915339
id: 9007203075656331