错误 ID: 25236
The tag data of a transport motor cannot be read. The motor ID is given in the event parameter: 0=A, 1=B, 2=Z1, 3=Z2.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | One of the sample handler cables is not properly connected. | 请与您的 Agilent 服务代表联系。 |
2 | One of the sample handler cables is damaged (corroded or chipped off). | 请与您的 Agilent 服务代表联系。 |
3 | 主板出现故障。 | 请与您的 Agilent 服务代表联系。 |
base-id: 3803988107
id: 9007203058729099