错误 ID: 25123
The autosampler failed to complete initialization correctly. The autosampler initialization procedure can recognize and move the injection valve to its reference positions in a predefined routine. During initialization, the processor monitors the position sensor, tag sensors, and actuator motor to check for correct movement. If one or more of the movements or the status information of the subparts is not read out successfully, the error message is generated.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Injection valve not installed correctly. | Check if the injection valve is installed correctly. |
2 | TAG and TAG reader not aligned correctly. | Check if the TAG or the TAG Reader are aligned correctly. |
3 | Electrical connection or components are defective. | 请与您的 Agilent 服务代表联系。 |