生物惰性模块仅使用生物惰性部件! |
对于生物兼容性模块,只能使用生物 / 生物兼容部件! 请勿与生物惰性部件混用。 |
适用于
|
所需工具
数量 | 部件号 | 描述 | |
---|---|---|---|
1 | ![]() | 8710-0510 | 开口扳手 1/4-5/16 英寸 |
所需部件
数量 | 部件号 | 描述 | |
---|---|---|---|
1 | ![]() | 5067-4232 | 2-position/6-port Injection Valve (VICI) 1300 bar (G7167B) |
1 | ![]() | 5067-6698 | 2-position/6-port RC Injection Valve 800 bar (G7167A) |
1 | ![]() | 5067-4260 | 2-position/8-port Injection Valve Dual Needle 1300 bar |
1 | ![]() | 5067-4263 | 2-position/6-port Injection Valve Bio-inert 600 bar for bio inert solution |
1 | ![]() | 5067-6739 | 2-position/6-port injection valve Bio 1300 bar (G7137A) |
前提条件
|
注意
Please bear in mind that depending on which valve you have installed the images may slightly differ from the actual item.
Disconnect the capillaries.
Turn the spanner nut counter clockwise until the injection valve head detaches from the hydraulic box (Do not use wrenches on the spanner nut).
Remove the spanner nut from the injection valve head.
将替换的进样阀头插入液压箱打开的传动槽中。旋转,直到替换进样阀头底部的接头与阀传动装置契合。
或
如果外销不能嵌合到外槽中,则必须转动阀头,直到感觉到两个销钉卡入槽内为止。现在,您在持续转动阀头直至销钉卡入槽内的过程中应该会感觉到来自阀驱动器的额外阻力。
注意
检查后侧的朝向。
验证阀标签的位置是否正确。
继续旋转,直到进样阀头中的计时销与外壳中的槽口对齐,然后将替换进样阀头压入执行器中。
Replace the spanner nut (1.) and tighten clockwise (2.) (Hand tighten only, do not use wrenches on the spanner nut).
Reconnect the capillaries