生物惰性模块仅使用生物惰性部件! |
对于生物兼容性模块,只能使用生物 / 生物兼容部件! 请勿与生物惰性部件混用。 |
适用于
|
所需工具
数量 | 部件号 | 描述 | |
---|---|---|---|
1 | ![]() | 8710-0510 | 开口扳手 1/4-5/16 英寸 |
所需部件
数量 | 部件号 | 描述 | |
---|---|---|---|
1 | ![]() | G4267-60300 | Sample Loop Flex 20 µL, right (red coded) |
1 | ![]() | G4267-60400 | Sample Loop Flex 40 µL, right (green coded) |
1 | ![]() | G4267-60500 | Sample Loop Flex 100 µL, right (blue coded) |
1 | ![]() | G7167-68500 | Sample Loop Cartridge 500 µL right |
1 | ![]() | G7167-68900 | Sample Loop Cartridge 900 µL right |
1 | ![]() | G5668-60500 | Bio-inert Sample Loop 100 µL (for G5668A) |
1 | ![]() | G7137-60300 | Sample Loop MP35N 20 µL, right (red/orange coded) |
1 | ![]() | G7137-60400 | Sample Loop MP35N 40 µL, right (green/orange coded) |
1 | ![]() | G7137-60500 | 样品定量环 MP35N 100 µL,右侧(蓝色/橙色编码) |
Further sample loops for the Dual-Needle option are available, see Sample Loops and Capillaries (Dual Needle) .
前提条件
|
警告
没有罩的进样针有造成人身伤害的危险
没有罩的进样针有对操作人员造成伤害的危险。
小心
样品定量环配置不匹配
系统损坏
注意
If you have changed the sample loop, verify that the correct sample loop is configured in the CDS (see Setting Up the Autosampler With OpenLab ChemStation ).
注意
For details on the setup of the dual-needle system, see Modify Capillaries .
小心
Blockages inside of the needle assembly union
Use a 1/4-inch wrench to tighten the fitting of the loop capillary.
Install the shorter capillary of the sample loop cartridge to the analytical head.
Pinch and reinsert the needle assembly and the connected sample loop capillary into the z- arm coupler.
注意
Check the tension of the loop capillary. This must be forced and guided to the hydraulic box to prevent it from being caught by the Z-drive.
In the Local Controller close Change needle /seat.
OR
In Agilent Lab Advisor software Change needle/loop. Click NEXT and wait until the needle is in the needle park station.
Click Back to leave the Maintenance window.
注意
If you need an autoreferencing step included, you must choose the change needle procedure.
注意
If you have changed the sample loop, verify that the correct sample loop is configured in the CDS (see Setting Up the Autosampler With OpenLab ChemStation ).