有必要对仪器进行定期检查,以确保其安全使用。Agilent 服务代表可以在合同的基础上进行这些定期检查。有关维护检查合同的信息,请联系您的 Agilent 代表。
The following pages describe maintenance (simple repairs) of the detector that can be carried out without opening the main cover.
Procedure | Typical Frequency | Notes |
---|---|---|
Flow cell exchange | If application requires a different flow cell type or if defective. | Complete Assembly A wavelength calibration check should be performed after replacement.
If the flow cell is removed and inserted, then a quick calibration check is performed. If this fails, you must do a wavelength recalibration, see Wavelength Verification and Calibration . |
Flow cell flushing | If flow cell is contaminated. |
|
Leak sensor drying | If leak has occurred. | Check for leaks. |
Leak handling System replacement | If broken or corroded. | Check for leaks. |