注意
New instruments are shipped with a flow cell installed and have a factory wavelength calibration.
If a flow cell is used other then the installed/shipped flow cell:
verify that the installed flow cell is fixed very tight,
perform a wavelength calibration using Glycogen (part of FLD Calibration Kit (G7121-68001)).
Preparation of the Glycogen Calibration Sample.
注意
The calibration solution needs to be prepared freshly. Once dissolved, Glycogen is only useable for 24 hours. Using water for calibration is not supported!
Preparation of the Flow Cell.
注意
Do not inject the calibration sample without the sample filter.
Additional hints for calibration:
Do not forget to reset the grating calibration in LabAdvisor.
Check that the cell has been fully inserted and fixed tightly.
Prepare a little more calibration solution, so all glassware can be rinsed thoroughly prior to using.
Never fill the syringe with the filter on.
Never empty the syringe completely.
With a 1mL glass syringe with Luer-lock, two to three injections of Glycogen might be required. Using a 5 mL glass syringe with Luer-Lock simplifies the procedure drastically lowers the risk of injecting air.
After the final injection, leave the syringe in the in-port and keep it in a horizontal position.
Wavelength Calibration.
Agilent Lab Advisor (preferred option): Under Instrument Control > Special Commands press the Reset grating calibration button, then go for Service & Diagnostics > Wavelength Calibration
Instant Pilot (G4208A): Maintenance > FLD > Calibration
注意
If the wavelength calibration process fails, refer to Wavelength Calibration Failed .
注意
Restore flow connections and flush with fresh water at 1.5 mL/min - 2 mL/min for at least 15-20 min. Prolonged flushing (e.g. overnight flushing) at low flow rates will not be sufficient.
Verification using the "Wavelength Accuracy Test".