错误 ID: 6703
This message may show up during a wavelength calibration.
If the expected deviation is larger than the specified wavelength accuracy, the message Wavelength Calibration Failed is displayed and the instrument stays in a Not Ready condition.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Flash lamp not ignited or position not correct. | 请与您的 Agilent 服务代表联系。 |
2 | Cell position not correct. | Check the cell position. |
3 | Solvent in the cell not clean or air bubble in the cell. | Flush the flow cell. |
4 | Monochromator assembly position not correct (after replacement). | 请与您的 Agilent 服务代表联系。 |