错误 ID: 6691
After exchanging the monochromator assemblies, the calibration factors should be reset to defaults values (a new FLM board comes with default values). In this case Wavelength Calibration Lost is displayed and the instrument stays in a Not Ready condition.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Reset of monochromator settings after exchange. | Perform a wavelength calibration. |
2 | Replacement of FLM board. | Perform a wavelength calibration. |
3 | Replacement of communication board | Perform a wavelength calibration. |
base-id: 3586212491
id: 9007202840953483