错误 ID: 7800, 7801, 7802, 7803, 7804, 7805, 7806, 7808, 7809, 7810, 7811, 7812, 7813, 7814, 7815, 7816
The motor test has failed.
Test 0 Failed: | Filter motor. |
Test 1Failed: | Grating motor. |
During the motor tests, the detector moves the motor to the end position while monitoring the end-position sensor. If the end position is not found, the error message is generated.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Motor is not connected. | 请与您的 Agilent 服务代表联系。 |
2 | Defective motor. | 请与您的 Agilent 服务代表联系。 |
3 | Defective/missing grating or filter. | 请与您的 Agilent 服务代表联系。 |
4 | Cable/connector defective. | 请与您的 Agilent 服务代表联系。 |