错误 ID: 1037
The calibration information needed for your detector to operate correctly has been lost.
During calibration of the detector the calibration values are stored in ROM. If no data is available in the spectrometer ROM, the error message is generated.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | The detector is new. | Recalibrate the detector. |
2 | The detector has been repaired. | 请与您的 Agilent 服务代表联系。 |