适用于
|
所需工具
数量 | 部件号 | 描述 | |
---|---|---|---|
1 | ![]()
|
| 扳手,1/4 英寸 for capillary connections |
1 | ![]() | 5043-0915 | 接头安装工具 |
1 | ![]()
|
| 4 mm 六角扳手 |
1 | ![]()
|
| 牙签 |
所需部件
数量 | 部件号 | 描述 | |
---|---|---|---|
1 | ![]()
|
| For parts, see Standard Flow Cell , Semi-Micro Flow Cell , Micro Flow Cell . |
前提条件
|
注意
The gaskets used in the standard and semi-micro/micro flow cell are different.
小心
样品降解和仪器污染
流路中的金属部件可能与样品中的生物分子发生反应,这会导致样品降解和污染。
Use a 4 mm hex key to unscrew the window assembly (1) and remove the gasket (6,7) from the cell body. Do not mix the gasket #6 and # 7.
Use a tooth pick to remove the quartz window from the window assembly.
注意
If the washers fall out of the window assembly, they must be inserted in the correct order with the PTFE ring to prevent any leaks from the flow cell window.
注意
Do not mix the gasket #6 and #7 (different hole diameter).
Using a 4-mm hex key, tighten the window screw hand tight plus a quarter turn.
Reconnect the capillaries, see 取出和安装流通池.
Perform a leak test.
Insert the flow cell.
Replace the front cover