The following pages describe maintenance (simple repairs) of the detector that can be carried out without opening the main cover.
Procedure | Typical Frequency | Notes |
---|---|---|
Cleaning of module |
If required. |
|
Deuterium lamp or tungsten lamp exchange |
If noise and/or drift exceeds your application limits or lamp does not ignite. |
An intensity test should be performed after replacement. |
Flow cell exchange |
If application requires a different flow cell type. |
A holmium or wavelength calibration test should be performed after replacement. |
Flow cell parts Cleaning or exchange |
If leaking or if intensity drops due to contaminated flow cell windows. |
A pressure tightness test should be done after repair. |
Holmium oxide filter Cleaning or exchange |
If contaminated. |
A holmium or wavelength calibration test should be performed after replacement. |
Leak sensor drying |
If leak has occurred. |
Check for leaks. |
Leak handling System replacement |
If broken or corroded. |
Check for leaks. |