This cuvette holder can be placed instead of a flow cell in the variable wavelength detector. Standard cuvettes with standards in it, for example, National Institute of Standards & Technology (NIST) holmium oxide solution standard, can be fixed in it.
This can be used for wavelength verifications.
适用于
|
所需部件
数量 | 部件号 | 描述 | |
---|---|---|---|
1 | ![]() | G1314-60200 | Cuvette Holder |
1 | ![]()
|
| Cuvette with the “standard”, e.g. NIST certified holmium oxide sample |
前提条件
|
Locate the cuvette holder on the desk.
Unscrew the bracket.
Insert the cuvette with the sample into the holder. The clear side of the cuvette must be visible.
Replace the bracket and fix the cuvette.
Install the cuvette holder in the instrument.
{注释者 , : ToDo: Condition: "not(($ProductGroup='Detector-VWD-BC' or $ProductGroup='G7114B'))" }Perform your Wavelength Verification/Calibration --- MISSING LINK --- to check the correct position of the cuvette holder.
Perform your Wavelength Verification/Calibration (see Wavelength Calibration) to check the correct position of the cuvette holder.
Perform your Wavelength Verification/Calibration (see --- MISSING LINK ---) to check the correct position of the cuvette holder.
base-id: 3284772363
id: 9007202539513355