有必要对仪器进行定期检查,以确保其安全使用。Agilent 服务代表可以在合同的基础上进行这些定期检查。有关维护检查合同的信息,请联系您的 Agilent 代表。
The following pages describe maintenance (simple repairs) of the detector that can be carried out without opening the main cover.
Procedures | Typical Frequency | Notes |
---|---|---|
Deuterium lamp exchange |
If noise and/or drift exceeds your application limits or lamp does not ignite. |
A wavelength calibration test and an intensity test should be performed after replacement. |
Flow cell exchange |
If application requires a different flow cell type. |
A wavelength calibration test should be performed after replacement. |
Cleaning flow cell parts cleaning or exchange |
If leaking or if intensity drops due to contaminated flow cell windows. |
A pressure tightness test should be done after repair. |
Leak sensor drying |
If leak has occurred. |
Check for leaks. |
Leak handling system replacement |
If broken or corroded. |
Check for leaks. |
base-id: 3284732811