错误 ID: 1072
This temperature sensor delivered a value outside the allowed range. The parameter of this event equals the measured temperature in 1/100 centigrade. As a result the temperature control is switched off.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | The temperature sensor is defect. | 请与您的 Agilent 服务代表联系。 |
2 | Detector is exposed to illegal ambient conditions. | Verify that the ambient conditions are within the allowed range. |