错误 ID: 205
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If the foam is removed, the fan is switched off, and the error message is generated.
The sensor on the main board detects when the top foam is in place. If the foam is removed during operation, the lamp and grating drive power is switched off, and the error message is generated.
The sensor on the main board detects when the top foam is in place. If the foam is removed, the pump is turned OFF within 4s for safety reasons, and the error message is generated. If the pump is started without cover, this error message is also generated and the pump is not turned on.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | The top foam was removed during operation. | 请与您的 Agilent 服务代表联系。 |
2 | The top foam was removed during operation or pump started without cover. | |
3 | Foam not activating the sensor. | 请与您的 Agilent 服务代表联系。 |
4 | Defective sensor or main board. | 请与您的 Agilent 服务代表联系。 |
5 | Rear of the module is exposed to strong direct sunlight. | Ensure that the rear of module is not directly exposed to strong sunlight. |