错误 ID: 24001
The valve drive was not able to operate the valve head correctly. Either due to mechanical reasons or the movement couldn’t be detected correctly.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Mechanical problems. Friction too high or blockages on the valve drive’s motor or on the valve head. | Check valve head for correct installation. Try to identify the source of trouble by installing a different valve head if possible. 请与您的 Agilent 服务代表联系。 |
2 | Defect Sensor on the Valve Drive Motor. | Check valve head for correct installation. Try to identify the source of trouble by installing a different valve head if possible. 请与您的 Agilent 服务代表联系。 |
base-id: 3181995531
id: 9007202436736523