错误 ID: 24000
During the initialization process the motor of the valve drive moves to some special positions depending on the installed valve head. A failure in this process means either that the movement couldn’t be performed properly or it was not noticed correctly by the sensor.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Mechanical problems. Friction too high or blockages on the valve drive’s motor or on the valve head. | Check valve head for correct installation. Try to identify the source of trouble by installing a different valve head if possible. 请与您的 Agilent 服务代表联系。 |
2 | Defect Sensor on the Valve Drive Motor. | Check valve head for correct installation. Try to identify the source of trouble by installing a different valve head if possible. 请与您的 Agilent 服务代表联系。 |
base-id: 3181957259