错误 ID: 32031
Switching of the valve failed.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Mechanical problems. Friction too high or blockages on the valve drive’s motor or on the valve head. | Check valve head for correct installation. Try to identify the source of trouble (valve head or drive) by installing a different valve head if possible. 请与您的 Agilent 服务代表联系。 |
2 | Defective sensor on the valve drive motor | Check valve head for correct installation. Try to identify the source of trouble (valve head or drive) by installing a different valve head if possible. 请与您的 Agilent 服务代表联系。 |
base-id: 3002720395