适用于
|
所需工具
数量 | 部件号 | 描述 | |
---|---|---|---|
1 | ![]()
|
| Tissue |
1 | ![]()
|
| Pipette |
1 | ![]()
|
| Wrench, 1/4 – 5/16 (for capillary connections) |
Remove the door.
Use a pipette and tissue to dry the leak sensor area.
Observe the capillary connections and the column switching valve for leaks and correct, if required.
Reinstall the door.
base-id: 3002045579