错误 ID: 80
The ambient-compensation sensor (NTC) on the mainboard in the module has failed (open circuit).
The resistance across the temperature compensation sensor (NTC) on the mainboard is dependent on ambient temperature. The change in resistance is used by the leak circuit to compensate for ambient temperature changes. If the resistance across the sensor increases above the upper limit, the error message is generated.
The ambient-compensation sensor (NTC) on the on/off switch board in the module has failed (open circuit).
The resistance across the temperature compensation sensor (NTC) on the on/off switch board is dependent on ambient temperature. The change in resistance is used by the leak circuit to compensate for ambient temperature changes. If the resistance across the sensor increases above the upper limit, the error message is generated.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Defective on/off switch assembly. | 请与您的 Agilent 服务代表联系。 |
2 | Loose connection between the on/off switch board and the mainboard. | 请与您的 Agilent 服务代表联系。 |
3 | Defective mainboard. | 请与您的 Agilent 服务代表联系。 |
base-id: 3002664971