错误 ID: 82
The leak sensor in the module has failed (short circuit).
The current through the leak sensor is dependent on temperature. A leak is detected when solvent cools the leak sensor, causing the leak sensor current to change within defined limits. If the current increases above the upper limit, the error message is generated.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Defective leak sensor. | 请与您的 Agilent 服务代表联系。 |
2 | Leak sensor incorrectly routed, being pinched by a metal component. | 请与您的 Agilent 服务代表联系。 |
3 | On/Off switch assembly defective. | 请与您的 Agilent 服务代表联系。 |
4 | Cable or contact problem. | 请与您的 Agilent 服务代表联系。 |