错误 ID: 83
The leak sensor in the module has failed (open circuit).
The current through the leak sensor is dependent on temperature. A leak is detected when solvent cools the leak sensor, causing the leak sensor current to change within defined limits. If the current falls outside the lower limit, the error message is generated.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Leak sensor not connected to the mainboard. | 请与您的 Agilent 服务代表联系。 |
2 | Leak sensor not connected to the on/off switch board. | 请与您的 Agilent 服务代表联系。 |
3 | Defective leak sensor. | 请与您的 Agilent 服务代表联系。 |
4 | Leak sensor incorrectly routed, being pinched by a metal component. | 请与您的 Agilent 服务代表联系。 |
5 | On/Off switch assembly defective. | 请与您的 Agilent 服务代表联系。 |
base-id: 3002558859
id: 54043198531004811