错误 ID: 22434
The current of the secondary pump drive is too high.
Parameter: 2 or 3 referring to pump drive
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Blockage of flow path between secondary pump head and pressure sensor, e.g. of the heat exchanger. | Check for blockages in the flow path. 请与您的 Agilent 服务代表联系。 |
2 | Secondary pump drive is defect. | 请与您的 Agilent 服务代表联系。 |