错误 ID: 22410, 29204
Drive is blocked mechanically, fails during initialization.
Parameter: 1 – 4 referring to pump drive
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Blockage of system before pressure sensor. | Check for blockage of e.g. outlet valve filter frit, purge valve, heat exchanger. |
2 | Drive encoder failed. | 请与您的 Agilent 服务代表联系。 |
base-id: 4202534155