错误 ID: 22407, 29205
The current consumption of the pump drive is too low.
Parameter: 1 – 4 referring to pump drive
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Drive motor defect. | 请与您的 Agilent 服务代表联系。 |
2 | Wrong/missing connection of pump drive to main board. | 请与您的 Agilent 服务代表联系。 |
base-id: 4202525195