错误 ID: 29256
The MCGV hardware check failed.
可能的原因
建议采取的措施
1
MCGV/4CGV may be disconnected.
Check MCGV/4CGV cable connection.
2
Valve defect.
请与您的 Agilent 服务代表联系。
Replace MCGV/4CGV.
base-id: 4133255435