错误 ID: 29221
The target pressure of the quaternary pump degasser has not been reached within the expected time.
Parameter: Pressure in mbar
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Condensation in degasser chamber due to temperature fluctuation. | Equilibrate and restart module. Use Evacuation Mode if necessary. |
2 | Degasser is defect. | 请与您的 Agilent 服务代表联系。 |
base-id: 4132515211
id: 4132515211