错误 ID: 2049
Valve Fuse 0: Channels A1 and A2
Valve Fuse 1: Channels B1 and B2
One of the solvent-selection valves in the module has drawn excessive current causing the selection-valve electronic fuse to open.
可能的原因 | 建议采取的措施 | |
---|---|---|
1 | Defective solvent selection valve. | Restart the pump. If the error message appears again, exchange the solvent selection valve. |
2 | Defective connection cable (front panel to main board). | 请与您的 Agilent 服务代表联系。 |
3 | 主板出现故障。 | 请与您的 Agilent 服务代表联系。 |
base-id: 3836597003